5/2
Low-flow cutoff 5.3
Display 5.4
Fct. 1.04 DISPLAY
Press
key.
DISP. FLOW = select unit for display of flowrate,
press
key
NO DISP. (no display)
m3/hr (cubic metres per hour)
Liter/Sec (litres per second)
US.Gal/min (US gallons per minute)
user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.15
PERCENT (percentage display)
BARGRAPH (numerical value and bar graph display in %)
Select with
key.
Press
key to transfer to subfunction “DISP. TOTAL”.
DISP. TOTAL = select unit for totalizer display,
press
key
NO DISP. (no display)
OFF (internal totalizer switched off)
• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (
ΣΣ
) • ALL (sequential)
Select with
key.
Transfer to totalizer unit setting using
key.
•m3 (cubic metres)
Liter (litres)
US.Gal (US gallons)
user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.12
Select with
key.
Transfer to totalizer format setting using
key.
Continuation see next page
Fct. 1.03 L.F.CUTOFF
Press
key.
Choice
OFF (fixed tripping point: ON = 0.1 % / OFF = 0.2 %)
PERCENT (variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)
Select with
key.
Transfer to number setting using
key (only if “PERCENT” selected).
1st number (cursor) flashes.
Setting the numerical value when “PERCENT” selected
01 to 19 (cutoff “on” value, left of hyphen)
02 to 20 (cutoff “off” value, right of hyphen)
Change flashing number (cursor) with the
key.
Shift cursor 1 place to right using
key.
Press
key to return to Fct. 1.03 L.F.CUTOFF.
Note: The cutoff “off” value must be greater than the cutoff “on” value.