5.8 Status outputs B1 and B2
NOTE: Check whether under Fct. 3.07 “HARDWARE” the output terminal “B1” and/or “B2” is
defined as status output B1 and/or B2, refer also to Sect. 2.1 and Sect. 5.16.
Fct. 1.06 and/or 1.07 ST
ATUS B1 and/or B2
Press key
→
.
Select function of status outputs,
press
→
key
• ALL ERROR (indicates all errors)
• FATAL.ERROR (indicates fatal errors only)
• OFF (switched off, no function)
•ON (indicates that flowmeter is operative)
• SIGN. I
V/R mode
• SIGN. P
• OVERFLOW I overranging I = I ONLY
• OVERFLOW P of outputs P = ALL
• EMPTY PIPE (option ”empty tube identification)
• AUTO RANGE (automatic range change) Settin
g range 5-80 PERCENT
(= ratio of upper to lower range, 1:20 to 1:1.25, value must be greater
than that of Fct. 1.03 “L.F.CUTOFF”, see also Sect. 5.18)
• TRIP POINT (define trip point) see also Sect. 5.17.
XXX – YYY
0 – 150% 0 – 150% N/O contact: XXX > YYY
N/C contact: XXX < YYY
Hysteresis: difference between XXX and YYY.
Press
↵
key to transfer to number setting, 1st digit (cursor) flashes.
Change flashing digit (cursor) with key
↑
. Use key
→
to shift cursor 1 place to right.
Press
↵
key to return to Fct. 1.06 and/or 1.07 STATUS B1 or B2.
Connection diagrams, see Sect. 2.6.
}
}
}
Dynamic response
of outputs, see Fct. 1.02, Sect. 5.2 “time constant”
Please refer to Sect. 3.2 ”factory settings“.
• Characteristics of status outputs Switch open Switch closed
OFF (switched off) no function
ON (e.g. operation indicator) power OFF power ON
SIGN I (F/R mode) Forward flow Reverse flow
SIGN P (F/R mode) Forward flow Reverse flow
TRIP POINT (limit switch) inactive active
AUTO RANGE
(automatic range change)
high range low range
OVERFLOW I (I overranged) current output OK current output overranged
OVERFLOW P (P overranged) pulse output OK pulse output overranged
ALL ERROR (all errors) errors no error
FATAL.ERROR (fatal errors only) errors no error
EMPTY PIPE (option) when measuring tube when measuring tube
is empty is full