4/7
F/R flow measurement
overranging
of outputs
}
dynamic response
of outputs, see Fct. 1.02
TIMECONST.:
I = I only
P = ALL
1.06 CONTROL B1 Control input B1 and B2 (see Fct. 3.07 HARDWARE)
1.07 CONTROL B2 OFF EXT.RANGE (external range change)
Setting range: 5 - 80 PERCENT (= ratio of lower to upper range from
1:20 to 1:1.25. Value must be greater than that of Fct. 1.03 L.F. CUTOFF).
Press
key to transfer to number setting.
OUTP.HOLD (hold value of outputs)
OUTP.ZERO (set outputs to
min.values“)
TOTAL. RESET (reset totalizers)
ERROR. RESET (delete error messages)
Press key
to return to
Fct. 1.06 or 1.07 CONTROL B1 or B2.
Fct. Text Description and settings
1.06 PULS B1 Pulse output B1 (see Fct. 3.07 HARDWARE)
FUNCT. P Select function for pulse output P
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key
to transfer to subfunction
“SELECT P”.
SELECT P Select pulse type
PULSE/VOL. (pulses per unit volume, flowrate)
PULSE/TIME (pulses per unit time for 100% flowrate)
Press key
to transfer to subfunction
“PULSWIDTH”.
PULSWIDTH Select pulse width
0.01 - 1.00 Sec (only for F
max
< 50 pulses/s)
AUTO (automatic = 50% of the period duration
of the 100% output frequency)
SYM. (symmetrical = pulse duty ratio 1:1 over total range)
Press key
to transfer to subfunction
“VALUE P”.
VALUE P Set pulse value per unit volume (appears only when
“PULSE/VOL.” set under “SELECT P” above)
xxxx PulS/m3 xxxx PulS/Liter xxxx PulS/US.Gal
xxxx PulS/
user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
Setting range
xxxx” is dependent on the pulse width and the
full-scale range: P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Press key
to return to
Fct.1.06
PULS. B1”.
VALUE P Set pulse value per unit time (appears only when
PULSE/TIME”
set under
SELECT P” above)
xxxx PulSe/Sec (=Hz) xxxx PulSe/min xxxx PulSe/hr
xxxx PulSe/user-defined unit, factory-set is “hr” or “day” (see Fct. 3.05)
Setting range
xxxx” is dependent on the pulse width, see above
Press key
to return to
Fct.1.06
PULS. B1”.
1.06 STATUS B1 Status output B1 and B2 (see Fct. 3.07 HARDWARE)
1.07 STATUS B2 ALL ERROR FATAL ERROR OFF ON
• SIGN. I
• SIGN. P
}
OVERFLOW I
OVERFLOW P
}
AUTO RANGE (automatic range change)
Setting range 5 - 80 PERCENT (= ratio of lower to upper range from
1:20 to 1:1.25. Value must be higher than that of Fct. 1.03 L.F. CUTOFF)
TRIP POINT: XXX - YYY XXX > YYY N/O contact
0 - 150% 0 - 150% XXX < YYY N/C contact
Hysteresis 1%
(difference between XXX value and YYY value)
EMPTY PIPE (signals that pipe is
empty”, only if option installed)
Press key
to transfer to number setting.
Press key
to return to
Fct. 1.06 or 1.07 STATUS B1
or
B2.