10/2
Error limits at reference conditions 10.3
Display, digital values, pulse output
F maximum error in % of measured value (not typical values)
v Flow velocity in m/s and ft/s
Reference conditions similar to EN 2
9104
Product water at 10 – 30°C/50 – 86°F
Electrical conductivity > 300 µS/cm
Power supply (rated voltage) U
N
(± 2%)
Ambient temperature 20 – 22°C/68 – 71.6°F
Warm-up time 60 min
Inlet/outlet runs 10
x
DN/2
x
DN (DN = meter size)
Primary head properly grounded and centered
Type/Meter size Maximum error in % of measured value (MV) ...
Curve
DN mm
DN 2.5 – 6(1)
DN 2.5 – 6(1)
≥ DN 10
≥ DN 10
inch
1
/10” –
1
/4”(1)
DN 2.5 – 6(1)
≥
3
/8”
≥ DN 10
v ≥ 1 m/s / ≥ 3 ft/s
≤ ± 0.5% of MV
DN 2.5 – 6(1)
≤ ± 0.3% of MV
≥ DN 10
v < 1 m/s / < 3 ft/s
≤ ± (0.4% of MV + 1 mm/s)
≤ ± (0.4% of MV + 0.04 inch/s
≤ ± (0.2% of MV + 1 mm/s)
≤ ± (0.2% of MV + 0.04 inch/s)
B
D
A
E
Current output same error limits as above, additionally ± 10 µA
Reproducibility and
repeatability 0.1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow
External influences t
ypical values maximum values
Ambient temperature
Pulse output < 0.003% of MV (2) 0.015% of MV (2)
Current output < 0.013% of MV (2) 0.025% of MV (2)
Power su
pply < 0.023% of MV 0.055% of MV (2) at 10% variation
Load
< 0.013% of MV 0.025% of MV (2) at max. permissible load,
see Sect. 10.4
(1) IFM 6080 K and IFS 6000 F (DN 2.5 – 4 and 1/10 – 1/6”) additional error ± 0,3% of MV
(2) All Krohne signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures
– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.
}
per 1 K / 1.8°F variation
0 1.6 3.2 6.4 32 36 40 v [ft/s]