Always switch off power source before opening the housing !
Re
quired measuring instruments and tools
- Special wrench to unscrew the covers, a crosstip screwdriver
- Ohmmeter with at least 6 V range
or
AC voltage/resistance measuring bridge.
Note: Accurate measurements in the area of the electrodes can only be obtained with an
AC voltage/resistance bridge. Also, the measured resistance is very heavily dependent on
the electrical conductivity of the liquid product.
Preparatory work
Switch off power source.
Unscrew covers from terminal and electronic
compartments using the special wrench.
If provided, remove the display unit. For this
purpose, detach the two screws R and fold
display unit to the side, see figure in Sect. 8.5.
Pull off the blue 9-pin plug from the amplifier
PCB, see figure in Sect. 8.9, field power
supply (Pin 7+8) and signal cables
(Pin 1, 2, 3, 4 + 5)
Completely fill the measuring tube of the
flowmeter with the process liquid.
7.5 Test of primary head
Action Typical result Incorrect result =
Resistance measurement flowmeter defective,
at the blue 9-pin plug return to factory for repair,
refer to last but one page !
1 Measure resistance 30 - 150 if lower:
between wires interwinding fault
7 and 8.
if higher:
wire break
2 Measure > 10 M if lower:
resistance between interwinding fault to PE
U-clamp terminal or FE.
in terminal compartment
(PE protective conductor
or FE functional ground)
and wires 7 and 8
3 Measure resistance 1 k- 1 M if lower: drain measuring
between wires 1 and 3 (see “Note” above). tube and repeat
and between 1 and 4 Both values should be measurement; if still
(same measuring lead approx. equal. too low, short-circuit in
always on wire 1 ! ) electrode wires.
if higher: break in
electrode wires or
electrodes
contaminated.
Values not equal:
break in electrode
wires or electrodes
contaminated.
Jacks X1 and X4 on amplifier PCB,
see Sect. 8.9
Blue 9-pin plug
(connection to primary head)
unassigned
X1
X4
7 8 1 2 3 4 5