7/6
Part 2 Signal converter IFC 090 D (D = display version) and
Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)
Group I Fault / Symptom Cause Remedial action
I 1 Receiver instrument Incorrect connection/polarity Connect properly
indicates “0”. see Sect. 2.3 + 2.7.
Receiver instrument or current Check output (see Sect. 7.2)
output defective. with new milliammeter:
T
est ok, check connection
cables and receiver instrument,
replace if necessary.
T
est faulty, current output
defective. Replace signal
converter (see Sect. 8.7)
or contact Krohne Service
Current output disabled, Activate under Fct. 1.05.
see Fct. 1.05
Short-circuit between current Check connection and cables,
output and pulse output. see Sect. 2.3 + 2.7.
Voltage between I+ and I⊥
approx. 15 V. Switch off device,
eliminate short-circuit,
and switch device on again.
I 2 Unsteady display Low electrical conductivity, Increase time constant,
high solids content, see Fct. 1.02, or contact
pulsating flow. Krohne Service.